Accuracy of Factory-Calibrated Continuous Glucose Monitors in Critically Ill Patients Receiving Intravenous Insulin: A Prospective Clinical Trial of Two Leading Systems.

J Diabetes Sci TechnolMay 8, 2025 (epub)
Clinical ResearchEndocrinologyInternal Medicine (General)Open access

Gautam Ramesh, Emily Kobayashi, Navyaa Sharma et al.

✦ AI-curated · Sources linked

In 30 seconds

This nonrandomized, open-label, clinically blinded prospective trial assessed the accuracy of two factory-calibrated continuous glucose monitors in 40 critically ill hospitalized patients receiving continuous intravenous insulin. CGM readings were compared with laboratory serum and point-of-care glucose measurements. Mean absolute relative difference was 22.7% for Dexcom G6 Pro versus laboratory values and 25.2% for FreeStyle Libre Pro versus laboratory values; accuracy varied considerably between patients, and both systems tended to underestimate glucose. The findings suggest that factory calibration alone may not ensure adequate accuracy for inpatient use in this population.

Key findings

  • Dexcom G6 Pro had a 22.7% MARD against laboratory glucose and 22.9% against point-of-care glucose.
  • FreeStyle Libre Pro had a 25.2% MARD against laboratory glucose and 27.0% against point-of-care glucose.
  • Both systems showed considerable inter-patient variability and tended to underestimate glucose relative to reference measurements.
  • Low reference glucose values were rarely overestimated by either sensor.

Why it matters

Glucose measurements guide insulin management in critically ill patients, so sensor inaccuracies may affect interpretation of glycaemic status. The observed variability and tendency toward underestimation indicate that performance in this setting cannot be assumed from factory calibration alone. These results help frame the need to evaluate CGM readings against established glucose measurements when considering inpatient use.

What to keep in mind

The trial was nonrandomized and included 40 patients; the abstract does not report clinical outcomes associated with CGM use.

Source

Published in J Diabetes Sci Technol. This summary was written by xxcode from the publication's abstract and metadata. It is not peer reviewed and is not a substitute for the original article. For clinical decisions, review the original publication.

ShareTelegramLinkedIn

AI-generated summaries may contain errors or omissions. Verify clinically important information with the original publication.

More in Endocrinology

Want this personalized?

Stop searching the literature. Choose what you follow and xxcode will build your personalized medical digest.

  • Your specialties
  • Your filters and thresholds
  • Clinical Research + Case Reports, tuned separately
  • Automatic weekly delivery
  • Audio and text

Prefer listening? Personalized audio digests are available with Pro.

Stay updated for free

Get the 3 most interesting publications in one specialty each week.

Weekly email. Unsubscribe anytime.