Impact of 4D-CT MIP reconstruction method on iGTV definition: A clinical case report.
Suhong Yu, Itai Pashtan
✦ AI-curated · Sources linked
Case in brief
A case during thoracic radiation therapy planning revealed a discrepancy in tumor extent due to the use of a phase-sorted 4D-CT maximum intensity projection (MIP) dataset, which truncated the inferior tumor boundary. A comparison with a cine-based MIP showed that the latter provided a more accurate representation of the tumor's motion envelope. This finding emphasized the need for thorough verification of MIP reconstruction methods in clinical practice.
Diagnostic / clinical pearl
Clinicians should verify the MIP reconstruction method used in 4D-CT imaging, as discrepancies can significantly affect internal gross tumor volume (iGTV) definition, especially in patients with irregular breathing patterns.
Why it matters
This case underscores the necessity for comprehensive quality assurance in motion management workflows to ensure accurate tumor delineation and effective treatment planning.